2 technical data, Technical data, Instrument components and specifications – Leica Biosystems ASP6025 User Manual
Page 20
18
Instructions for Use V 2.2 RevI – 08/2013
3.
Instrument Components and Specifications
3.2 Technical Data
Nominal supply voltages:
Two factory-preset voltages (not user-adjustable):
120 V + 10 %
or
230 V + 10 %
Nominal frequency:
50 to 60 Hz
Main fuses:
2 melting fuses, 32 x 6.3 mm, UL-approved
• for 120 V; T 15 A 125 VAC
• for 230 V; T 10 A 250 VAC
Nominal power:
1700 VA for both versions
Dimensions, (L x W x H), in mm:
680 x 750 x 1500
Empty weight, unpacked:
max. 210 kg
Weight, packed:
max. 235 kg
Operating temperature range:
+15 °C to +40 °C
Storage temperature range:
+5 °C to +50 °C
Relative humidity:
10 % to max. 80 % non-condensing
IEC 1010 classification:
Protection class 1
Pollution degree 2
Overvoltage installation category II:
Ambient pressure:
740 to 1100 hPa
Operating elevation:
up to a max. of 2000 m above sea level
A-weighted noise level:
< 70 dB (A)
Local/remote alarm relay:
30 V DC, max. 1 A
120 V AC, max. 0.25 A
2 terminals for 6.3 mm stereo jack.
Each with isolated switching contact
(operable both as normally-open and
normally-closed circuit)
Paraffin containers
Paraffin melting station:
1
Number of paraffin baths:
3
Capacity:
4.9 l per paraffin bath
Melting time:
in paraffin station: max. 6.0 h, in paraffin bath max. 12.0 h
Temperature:
50 to 65 °C (+ 6 K - 0 K)