18

Instructions for Use V 2.2 RevI – 08/2013

3.

Instrument Components and Specifications

3.2 Technical Data

Nominal supply voltages:

Two factory-preset voltages (not user-adjustable):

120 V + 10 %

or

230 V + 10 %

Nominal frequency:

50 to 60 Hz

Main fuses:

2 melting fuses, 32 x 6.3 mm, UL-approved

• for 120 V; T 15 A 125 VAC

• for 230 V; T 10 A 250 VAC

Nominal power:

1700 VA for both versions

Dimensions, (L x W x H), in mm:

680 x 750 x 1500

Empty weight, unpacked:

max. 210 kg

Weight, packed:

max. 235 kg

Operating temperature range:

+15 °C to +40 °C

Storage temperature range:

+5 °C to +50 °C

Relative humidity:

10 % to max. 80 % non-condensing

IEC 1010 classification:

Protection class 1

Pollution degree 2

Overvoltage installation category II:

Ambient pressure:

740 to 1100 hPa

Operating elevation:

up to a max. of 2000 m above sea level

A-weighted noise level:

< 70 dB (A)

Local/remote alarm relay:

30 V DC, max. 1 A

120 V AC, max. 0.25 A

2 terminals for 6.3 mm stereo jack.

Each with isolated switching contact
(operable both as normally-open and
normally-closed circuit)

Paraffin containers
Paraffin melting station:

1

Number of paraffin baths:

3

Capacity:

4.9 l per paraffin bath

Melting time:

in paraffin station: max. 6.0 h, in paraffin bath max. 12.0 h

Temperature:

50 to 65 °C (+ 6 K - 0 K)

Advertising