• If the paraffin has been transferred out of the paraffin station

and into a paraffin bath using the SMART SCREEN menu,
the paraffin station is automatically set to "Empty" in the
Reagent Status.

Fig. 173

• Therefore, the paraffin station must be set to "Full" in the

reagent status (

Fig. 173

) menu to activate heating and melt-

ing after being refilled with paraffin pellets.

Filling a paraffin bath

(continued)

• To do so, go to the reagent status, select the paraffin station

and press the SET AS FULL button.

5. Operation

When the paraffin station is closed, a message appears explaining that the MELT button can be pressed to set
the paraffin station to FULL in the reagent status and to recalculate the melting time.

Caution!

If the paraffin baths are set to "Empty" in the Reagent status (

Fig. 173

), heating is also deactivated at the same

time. This means that any paraffin residues can solidify and the selected bath is thus not usable.