Initial Operation

Unpacking the instrument

(continued)

Removing the transport anchors

• After installing the monitor, you have to remove all transport

anchors (adhesive tape and foam parts).

• First, carefully remove all blue strips of adhesive tape (105)

that secure the doors and the drawer of the instrument.

105

Fig. 17

9

Fig. 18

• In the retort, you also have to remove the strip of adhesive

tape (105) that secures the cover of the level sensors
(

Fig. 19

).

• Then, open the drawer (9) and remove all foam parts there

(

Fig. 18

). Also remove the six dark gray molded parts (106)

from the necks of the RTU-bottles.

Fig. 19

106

19

19

105

9

• Remove the adhesive tape (105) from the lid of the retort

(19) (

Fig. 17

).

105