Initial operation – Leica Biosystems ASP6025 User Manual

Page 50

Advertising
background image

48

Instructions for Use V 2.2 RevI – 08/2013

4.

Initial Operation

Pull-out drawer with RTU-bottles

(continued)

Fig. 47

• The bottom right of the drawer has a spigot (64,

Fig. 47

),

for draining spilled or overflowing reagents neatly. Then,
the drawer should be cleaned.

Spigot closed

Spigot open

64

64

Fig. 46

10

• Seven RTU-bottles (10) are supplied for the drawer with the instrument (

Fig. 46

)

- 6x for tissue infiltration,
- 1x for the cleaning process (extended cleaning program by detergent).
Each container has a screw plug.
• Each container has a max. capacity of 5 l.

Fig. 47a detail

Only RTU-bottles from Leica are approved for this instrument. However, if dif-

ferent bottles are used, the customer must INDEPENDENTLY check that they are

suitable for such use.

(For information about temperature and dimension requirements, see

page 49 -

Warning message.)

Important!

Always keep the spigot (Fig. 64 detail) closed.

The spigot (

Fig. 47

a

detail

, item 64) may be

opened for cleaning purposes only, as other-

wise flammable reagents could end up on the

heater of the paraffin baths and ignite.