Initial operation – Leica Biosystems ASP6025 User Manual
Page 50
48
Instructions for Use V 2.2 RevI – 08/2013
4.
Initial Operation
Pull-out drawer with RTU-bottles
(continued)
Fig. 47
• The bottom right of the drawer has a spigot (64,
Fig. 47
),
for draining spilled or overflowing reagents neatly. Then,
the drawer should be cleaned.
Spigot closed
Spigot open
64
64
Fig. 46
10
• Seven RTU-bottles (10) are supplied for the drawer with the instrument (
Fig. 46
)
- 6x for tissue infiltration,
- 1x for the cleaning process (extended cleaning program by detergent).
Each container has a screw plug.
• Each container has a max. capacity of 5 l.
Fig. 47a detail
Only RTU-bottles from Leica are approved for this instrument. However, if dif-
ferent bottles are used, the customer must INDEPENDENTLY check that they are
suitable for such use.
(For information about temperature and dimension requirements, see
page 49 -
Warning message.)
Important!
Always keep the spigot (Fig. 64 detail) closed.
The spigot (
Fig. 47
a
detail
, item 64) may be
opened for cleaning purposes only, as other-
wise flammable reagents could end up on the
heater of the paraffin baths and ignite.