7.2 Malfunctions in the program sequence

Maintaining the specimen in a safe state is the primary concern during a malfunction.
A safe specimen state is initiated following any power outage or malfunction where the infiltration
protocol cannot be automatically finished. An infiltration protocol that has been started cannot be
finished automatically leads to a safe specimen state being initiated.
Example: a reagent station is under filled and no alternative reagent is available.
Likewise, defective instrument components such as valves or a pump may potentially result in a
safe specimen state being initiated so the tissue processing is not endangered.

The safe specimen state of the instrument is indicated on the

SMART screen with an icon (109) immediately to the left of
the retort

(Fig. 240

).

In addition, a corresponding entry in the Run Log is generated.

Fig. 240

109

7. Troubleshooting