89

Leica ASP6025

This function is used to transfer instrument data from a USB stick
into an ASP6025. The following data is saved:
• all programs,
• the reagent list,
• the instrument status,
• the instrument configuration,
• all log files,
• the instrument installation parameters.

Saving or loading data

(continued)

• The instrument software will check that the instrument

data on the stick is actually complete. If it is not, the loading
operation will be terminated.

Fig. 119

Loading data from a USB stick

To write the instrument data from USB stick back to the instru-
ment, proceed as follows:
• Insert a USB stick in one of the USB ports on the instrument

and press the LOAD button.

• Press the ENTRIES and LOAD FROM USB buttons to open

the window of the same name.

• Once all data has been loaded successfully, a message

confirming this appears on the screen and prompts the
user to check that the reagents have been assigned correctly
(

Fig. 120

).

• Finally, successful completion of the entire operation is also

indicated in a message.

Fig. 120

5. Operation

Since there is already instrument data on the

ASP6025, this will be overwritten and thus

deleted. This will be indicated in a warning

message (

Fig. 119

) and must be confirmed

with YES. Pressing the NO button terminates

the process.

Advertising