Leica ASP6025

5.4.2 Refilling paraffin

5. Operation

There are two marks in the interior of the station that display the minimum fill level when filled
with pellets or liquid paraffin respectively (

Fig. 170

). The level must not fall below these marks.

The paraffin station can be filled with paraffin pellets, paraffin blocks or liquid paraffin.
If filled with pellets, the melting time is approx. 6.0 h.

Filling a paraffin bath

• In the smart screen, select an empty paraffin bath and the

full paraffin station. Then press the STATION REFILL button.

• Confirm the prompt (

Fig. 1

71

)

with YES.

• The liquid paraffin is then pumped from the paraffin station

to the retort.

• Then, the selected paraffin bath is filled from the retort.
• Finally, the retort has to be cleaned.

Fig. 171

Fig. 172

The paraffin station

Fig. 170

+

If a paraffin bath is filled with pellets, the melting time will be
approximately 720 min (12 h).
• Fill the paraffin bath to the brim with pellets, and after about

6 h top up with pellets again.

• Then set the melting time in the SERVICE FUNCTIONS menu

(see also

Chapter 5.1.5

).

Filling manually

Filling from the paraffin station

The initial filling of all three paraffin baths should have used paraffin that has been melted earlier, e.g. using

an external paraffin oven or other processor, so that the instrument can be started up as quickly as possible.

Then, use the paraffin station for filling.

If the last reagent in the retort is not com-

patible with paraffin, the retort must first be

cleaned using a cleaning program.