4.

Initial Operation

Setting up the ramp

Unpacking the instrument

(continued)

Fig. 8

93

92

94

91

91

• Pull out the two rails (91)

for the ramp beneath the
instrument by pulling
them forwards (

Fig. 8

).

91

• Hook these rails on the left

and right with the panel
(92) into the groove (93)
of the pallet so that they
are flush with the wooden
board (94) that holds the
instrument (

Fig. 9

).

95

95

Fig. 10

Fig. 9

• While doing so, make sure that the plate (92) is between the

two screws (95) in the groove. These screws prevent the rail
from being able to shift laterally.

97

96

• To do so, the lever must be folded upwards.

• Now release the brake lever
(97) on the two front transport

castors (96) so that the instru-

ment can be moved (

Fig. 10

).