Leica ASP6025

6. Cleaning and Maintenance

6.2.2 Periodic cleaning and maintenance

Cleaning the system bottles
• Empty and clean the system bottles.

Using a bottle brush and a laboratory detergent in warm water.

Care of reagent and condensate bottle seals
• To ensure easy removal of the reagent and condensate bottles, lubricate the O-ring seals

(65.

1

,

Fig. 235

and

236

) on the plug-in nozzles with the O-ring lubricant supplied.

Connection

port with

delivery hose

65.1

Fig. 236

Connection

port

Condensate bottle

System bottle

Fig. 235

65

35

35

65.1

65

The level can be identified by simply looking through the system bottles if these

are cleaned regularly to remove clinging contamination.

Never clean system bottles in an automatic dishwasher.

The reagent containers are NOT dishwasher-proof!

The connection manifolds must NEVER be mixed up. The connection manifold

with the delivery hose must ALWAYS be inserted in the white system bottle.

Tighten the retaining ring (35) so that the bottle is perfectly sealed.