28

Instructions for Use V 2.2 RevI – 08/2013

4.

Initial Operation

4.1 Unpacking the instrument

• Bring the transport crate (80) in which the

instrument is packaged as close as possible to
the final installation location.

• There, first remove the straps (82), then the

cover (81).

Fig. 5

84

80

81

82

83

82

Opening the package (

Fig. 5

)

1 m

1 m

1 m

3 m

ASP6025

Box

Front

Sufficient surface area is required for unpacking the
instrument.
The distance to the nearest wall must be at least
1 m on the side and rear.
On the front, the distance must be at least 3 m,
as the ASP6025 is rolled from the pallet in this
direction.
The room height must be at least 2.5 m, as the
packaging must be removed by pulling it upwards.

Fig. 4

85

• Important!
• At least TWO persons are required when lifting or carrying the instrument.
• The packaging has two indicators (83, 84,

Fig. 5

) that indicate improper transport. When the instrument is

delivered, check this first. If one of the indicators is colored red, the package was not handled as prescribed.

• In this case, please mark the shipping documents accordingly and check the shipment for damage!