167

Leica ASP6025

7. Troubleshooting

• If no infiltration program is running when a power failure is detected, no further actions / backup measures are taken -

the instrument shuts down automatically.

When a program is running:
• The program that is running is stopped, regardless of the current status of processing.

Fig. 239

• If a reagent is being used for which no safe reagent is in the

instrument when the program starts, a warning message
appears (

Fig. 239

). However, the program launch is NOT

aborted.

• If a safe reagent is in the retort, the reagent can be kept in

the event of power failure, though the retort may have to be
filled to a high enough level so that no tissue sample is dry.

This operation is automatically controlled by the instrument

and cannot be influenced by the operator.

• If no safe reagent is in the retort, depending on the cur-

rent status of processing the program that is in progress, a
"decision" is made about how to proceed.

Current step for power failure

Safeguard in safe reagent

Formalin (fixation)

Formalin

Ethanol (dehydration, diluted)

Formalin

Ethanol (absolute)

Formalin

Xylene (intermedium)

Xylene (heating off)

Isopropanol (intermedium)

Formalin

Paraffin

Paraffin (heating off)

ParaLast ParaLast

To conserve the battery, all heaters, the agitator and the monitor are switched off directly.