Instructions for Use V 2.2 RevI – 08/2013

5. Operation

Draining a paraffin bath

(continued)

Fig. 177

• The following security prompt must be confirmed in order

to continue (

Fig. 178

).

Fig. 178

• If necessary, the retort must be heated to the necessary

temperature so that the paraffin remains fluid. Then, the
paraffin is pumped from the selected paraffin bath into the
retort.

• When the bath is emptied, the paraffin is automatically

pumped from the retort into the external container.

• After this operation is complete, the paraffin residues are

blown out of the external tube and the hose. The correspond-
ing message must be confirmed with OK.

Fig. 179

+

Fig. 176

• Select the paraffin tank to be emptied and the receptacle for

the used paraffin (external station) (

Fig. 176

).

• Then press STATION DRAIN.

If the last reagent in the retort is not compatible with paraffin, the retort must first be cleaned using a cleaning

program.