40

Instructions for Use V 2.2 RevI – 08/2013

4.

Initial Operation

Fig. 30

41

42

Attach the magnetic stirrer (41) to the axis

on the floor of the retort with the little hole

facing upwards.

Insert the sieve (42) into the drain opening on

the floor of the retort.

Every time a program is started (except in CONCENTRATION
mode, see

Chapter 5.1.2

), a message is first displayed in which

the start must be confirmed.
Depending on how many baskets the retort is loaded with, the
fill volume can be selected in the dialog box by pressing the
corresponding button (

Fig. 31

).

• For this purpose, there are three level sensors (40.

1

- 40.

3

)

(

Fig. 31

) located behind the cover (40) in the back panel of

the retort.

• The uppermost sensor (40.

4

) prevents the retort from being

overfilled or overflowing.

40.

1

40.

2

40.

3

40.

4

The retort

(continued)

Filling the retort up to sensor 40.

1

Filling the retort up to sensor 40.

2

Filling the retort up to sensor 40.

3

Level sensors (

at left in Fig. 31

)

Fig. 31

Selection

buttons

Fill level

sensors

40

It is imperative to insert the mag-

netic stirrer before an infiltration

program is started!