Leica ASP6025

Pressing the RUN OPTIONS button opens
the window with the same name.

Fig. 192

Starting a program

(continued)

PRE-TEST activated:
After the program is started, the retort is filled and drained

with the first reagent in the infiltration program to test
whether all lines and valves are unobstructed.

PRINTER, if activated:
At the end of the program, a complete run log covering the

entire program execution is stored in the form of a PDF file.

Fig. 191

If you start a pro-
gram for which TIME

REQUEST was selected

as the end time, an entry window (

Fig. 191

) opens in which you

can select the end time.

This function (

Fig. 192

) shows the options with which the

current program was started.

RUN OPTIONS

To ensure that the program ends on the speci-

fied date and time, the ASP6025 extends the

duration of the programmed delay step cor-

respondingly.

When working with fixed end times, you

should always check that the instrument time

and date are current.

The options set apply for ALL programs!

All run options can be enabled or disabled according to requirements while the program is running.