5 ending a program, Operation – Leica Biosystems ASP6025 User Manual
Page 134
Advertising
132
Instructions for Use V 2.2 RevI – 08/2013
5. Operation
5.5.5 Ending a program
After fully draining the paraffin, an automatic drain time of 2 minutes is implemented in order to optimize the
use of cleaning reagents.
Once an infiltration program has ended, the retort can be drained basket by basket so that the cassette baskets can be
removed without coming into contact with hot paraffin.
Fig. 193