133

Leica ASP6025

5. Operation

5.6 Permanently installed infiltration programs

• The replacement cycle is defined by the ethanol concentration threshold value for the first ethanol measured by the

density sensor.

The ethanol threshold value has been set to 55 % by the manufacturer and can be adjusted in 1 % increments between

50 and 60 %.

• The ethanol concentrations shown are the initial concentrations that apply for installation of the tissue processor.

Reagent positioning by autorotation programs

5.6.1 Programs with automatic ethanol rotation

All of the permanently installed infiltration protocols described in this chapter have to be validated before use,

i.e. tissue processing using patient tissue for diagnostics, by the laboratory itself based on tissues that are not

intended for diagnostics and in accordance with local or regional accreditation requirements.

Autorotation programs cannot be renamed, edited or deleted by the user.

To start an autorotation program, it must be saved in the list of "FAVORITES". See also

Chapter 5.5.3

Bottle

Reagent

S1

70 % ethanol

S2

80 % ethanol

S3

95 % ethanol

S4

100 % ethanol

S5

100 % ethanol

S6

100 % ethanol

S7

Xylene

S8

Xylene

S9

Xylene

D1

Formalin

D2

Water

D3

Replacing xylene

D4

Replacing ethanol -> 100 %

D5

Cleaning xylene

D6

Cleaning alcohol