6 permanently installed infiltration programs, 1 programs with automatic ethanol rotation, Permanently installed infiltration programs – Leica Biosystems ASP6025 User Manual
Page 135: Operation
133
Leica ASP6025
5. Operation
5.6 Permanently installed infiltration programs
• The replacement cycle is defined by the ethanol concentration threshold value for the first ethanol measured by the
density sensor.
The ethanol threshold value has been set to 55 % by the manufacturer and can be adjusted in 1 % increments between
50 and 60 %.
• The ethanol concentrations shown are the initial concentrations that apply for installation of the tissue processor.
Reagent positioning by autorotation programs
5.6.1 Programs with automatic ethanol rotation
All of the permanently installed infiltration protocols described in this chapter have to be validated before use,
i.e. tissue processing using patient tissue for diagnostics, by the laboratory itself based on tissues that are not
intended for diagnostics and in accordance with local or regional accreditation requirements.
Autorotation programs cannot be renamed, edited or deleted by the user.
To start an autorotation program, it must be saved in the list of "FAVORITES". See also
Chapter 5.5.3
Bottle
Reagent
S1
70 % ethanol
S2
80 % ethanol
S3
95 % ethanol
S4
100 % ethanol
S5
100 % ethanol
S6
100 % ethanol
S7
Xylene
S8
Xylene
S9
Xylene
D1
Formalin
D2
Water
D3
Replacing xylene
D4
Replacing ethanol -> 100 %
D5
Cleaning xylene
D6
Cleaning alcohol