118

Instructions for Use V 2.2 RevI – 08/2013

Fig. 169

5. Operation

• Remove the system bottle (15) from the reagent cabinet and

unscrew the container cover (15.

1

).

• Drain the used reagent into a bulk container and then refill

the system bottle.

The fill levels for filling with 3.8 l and 5.0 l are embossed

on the front of each container.

Use a funnel for clean filling.
• Place the filled system bottle back in the reagent cabinet.
• When inserting the system bottle, always make sure that the

retaining ring (35) is securely tightened and the connection
spout (65) clicks into the back panel correctly.

• Then, in the SETTINGS window press the REAGENT STATUS

button and call up the table of the same name (

Fig. 169

).

• Select the corresponding line in the table and set the data

to zero using the DATA TO ZERO button (

Fig. 169

).

Manually changing the reagents

15

15.

1

35

Fig. 168

65

Caution!

Never replace reagents or fill empty system

bottles or RTU-bottles while a process is

running.

This can result in severe damage to the instru-

ment.

Manual filling has been described previously

with reference to the system bottles.

Follow the exact same procedure for the

RTU-bottles from the drawer.