69

Leica ASP6025

5. Operation

PROMPT FOR NUM CASSETTES:
Here, too, there are two options:

"Enabled" or "Disabled"

The current setting is likewise displayed next to the button.

Fig. 83

System settings

(continued)

ENABLED mode
When a program is started, you must enter the number of
cassettes that are inserted into the retort before an infiltration
program can be started.
This is useful if threshold values have been set.

Fig. 84

• In order to disable the PROMPT FOR NUM CASSETTES

function, press the button of the same name.

DISABLED mode
The number of cassettes inserted into the retort is not queried
prior to the program start.

If values for a reagent that is used in a pro-

gram are entered in the CASSETTES UNTIL

CHANGE column in the SET UP REAGENTS AND

WARNING THRESHOLDS, the mode is auto-

matically set to ENABLED.

This setting cannot be disabled afterwards.

If values have been entered there, a warning

message appears (Fig. 84).

ENABLED mode also means that the values

specified in this table will be observed.

If a warning threshold is reached or exceeded,

a message to that effect is displayed before

the program is started.

A warning message is also output at the end

of an infiltration program if the threshold

values have been exceeded.

Advertising