Operation – Leica Biosystems ASP6025 User Manual
Page 119

117
Leica ASP6025
• The entire operation is displayed graphically on the screen
(SMART FUNCTIONS view) (
Fig. 165
).
• The reagent is pumped out of the external container and
initially into the retort. Then, the selected system bottle is
filled.
Fig. 165
Fig. 166
Fig. 167
• When the selected station is filled with reagent, a message
window appears (
Fig. 16
6), with a message to the effect that
the hose for remote filling/draining is being blown out with air.
• Therefore leave the hose in the remote station until the
process has ended, or insert it into another suitable container.
• After the rinsing has been completed, a message appears
(
Fig. 167
), informing you that remote filling has been com-
pleted. To end the process, press OK.
Draining/filling reagents
(continued)
5. Operation