4.

Initial Operation

Fig. 26

4.4 Connecting the power supply

37

39

36

Electrical ports on the rear panel of the instrument.

Type plate for 120 V voltage variant.

Fig. 25

32

33

Adhesive label with fuse ratings

Check the nameplate (

Fig. 25

) on the rear of the

instrument to ensure that the instrument deliv-

ered is designed for the required voltage.

Severe damage may occur if the instrument

is connected to a power supply voltage other

than that to which it was originally set.

The power supply voltage for the instrument

is factory preset and CANNOT be altered by the

user.

Caution!

Observe the following instructions carefully to prevent damage to the instrument:

The 120 V voltage version of the instrument (REF 14 0495 43268) requires an electrical power supply with fuse

protection of at least 20 A.

The instrument MUST be connected to a grounded power socket. The power plug must be easily accessible so

that it can be pulled out.

The instrument is supplied with a set of different power cords. Use only the power cable that matches the local

power supply (outlet).

Do not use an extension cable!