Operation – Leica Biosystems ASP6025 User Manual

Page 117

Advertising
background image

115

Leica ASP6025

• On the screen (

Fig. 15

6), select the system bottle or

RTU-bottle that is to be drained and then select the recep-
tacle (external station).

Then press STATION DRAIN.

Select station

Press button

Draining/filling reagents

(continued)

Fig. 158

Fig. 159

Fig. 161

• A message window on the screen shows that the process has

been completed (

Fig. 161

). To continue working, confirm this

with OK.

• If the same reagent will not be used in the next operating

step, the retort may have to be cleaned with a cleaning
program.

• A prompt appears for carrying out the necessary preparations

(

Fig. 15

9). This must be confirmed with YES.

• The reagent is pumped out of the RTU-bottle into the retort.
• Once this task has been completed, the retort is drained into

the external station (79

in

Figs. 156

and

157

).

• The entire process is displayed graphically on the screen

(SMART SCREEN view) (

Fig. 153,

right).

Fig. 160

5. Operation

+