7. Troubleshooting

7.1 Troubleshooting

When a problem occurs in the Leica ASP6025, take the following steps to diagnose the problem:

• Check for error messages indicating the nature of the problem.
• If an error message is displayed, press the Help button on the message for any information

relating to the cause.

• Check the Run Log for any messages that indicate the nature of the failure. If the failure

occurred during a program run, check the Run Log for an indication of whether the failure
occurred during the fill, processing or drain cycles, and the station in use at the time.

7.1.1 Power failures

If there is evidence of power to the instrument:

• Verify that the power plug is plugged into the socket and that the socket is switched ON.
• Verify that the power switches on the rear of the instrument (adjacent to the power lead), and

on the side of the instrument are both switched ON.

• Error messages indicating that there has been a power failure mean that power has been lost

to the heaters and the monitor, but not to the main control electronics. Perform test procedure
as described in "Power failure" (

Chapter 7.1.2

).

Use the functions in the SMART SCREEN menu to test operating steps individu-

ally, such as the fill, drain, pressure, and vacuum steps.

If you cannot identify the problem by carrying out the above listed steps, call

Leica Technical Service.