Operation – Leica Biosystems ASP6025 User Manual

Page 99

Advertising
background image

97

Leica ASP6025

Reagent warning messages
If the RMS is active, warning messages are output if reagents are used for too
long.
The following indicates that the warning threshold values have been exceeded:
• Exclamation mark on the station system or RTU-bottle in the SMART screen

(

Fig. 13

5).

The RMS system – adjusting warning threshold values

(continued)

• When a warning threshold value is exceeded, a warning message is also

displayed at the end of a program (after the cleaning) (

Fig. 13

7).

From there, you can press YES to go directly to the REAGENT STATUS window.

Fig. 135

Fig. 136

Fig. 137

Fig. 138

• A warning message when starting a protocol (

Fig. 13

6).

The operator is notified that the reagent has run out and can replace this

reagent before the program starts.

However, it is also possible to start the program despite this.

• The reagent status highlights the exceeded values in red in the table (

Fig. 13

8).

Displayed if:
• The number of processed cassettes is exceeded.
• The number of specified cycles is exceeded.
• The days until changing the reagent are exceeded.

5. Operation