5. Operation

System settings

(continued)

The RUN OPTIONS function displays the options with which a
program is started.
The green check mark indicates that the function has been
enabled.

Pre-test:
• If this option is enabled, a retort filling and draining test

is carried out with the first reagent in the program before
the program is started. This is to ensure that the system is
functioning properly.

The configured option of a respective program always applies to

ALL infiltration programs!

Printer:
if this function is selected, a program report is generated after the infiltration is complete and can be saved as a PDF file.
The report contains the following data on the program that has run:
• The program name
• The program number
• The infiltration program that was executed
• Run log events that occurred during the program execution.
The program reports can be saved to a USB stick using the SAVE PDF PRINTOUTS in the ENTRIES menu.

Fig. 94

Select RUN OPTIONS

If a program has been selected or is already running, the run options can also be changed immediately at

any time in operator mode by pressing the RUN OPTIONS button in the graphic program window.

Make the change and confirm it with OK in the window that opens.

This change applies to all subsequent programs.