6.

Cleaning and Maintenance

6.2 General instrument cleaning

• Wipe the paraffin baths and lids clean.
The lid can be removed for cleaning purposes.
• Wipe the paraffin station and lid clean.
• If they are very dirty, remove the wax sieves from the wax baths and the paraffin station. Clean,

dry, and reinsert them.

Instrument exterior
Clean the instrument exterior as necessary. Wipe with a damp cloth moistened with mild detergent
and dry.

Checking the drip tray

Fig. 224

16

The collecting tray (16) must be inspected regularly for signs
of leaking reagents (

Fig. 224

). It is recommended to do this

after the infiltration and cleaning program has been completed,
before any leaking reagent has a chance to evaporate.
• To do so, pull the tray out by the handle (

Fig. 224)

.

Paraffin containers

Work carefully and wear gloves!

Be careful as the walls of the paraffin baths are very hot and may cause burns!

Caution!

Do not use solvents on painted surfaces, warning labels or the touchscreen.

CAUTION!

Some of the reagents used in the Leica

ASP6025 are hazardous to health. Therefore

always wear gloves and safety goggles when

handling them.