Leica ASP6025

6. Cleaning and Maintenance

Running cleaning programs

(continued)

Fig. 213

Cleaning after completion of an infiltration program

If desired, you can select an additional water step for each retort
cleaning program (EXCEPT the full cleaning with detergents).
(See also

Chapter 6.1.3

).

Fig. 214

• To select the water step, press the WATER YES/NO button so

that a green check mark appears (

Fig. 212/213

). The water

step is always carried out as the third step in the cleaning
program, that is to say it is not run in addition to the cleaning
program shown in

Fig. 208

.

The three cleaning programs are offered au-

tomatically when an infiltration program has

been completed (

Fig. 211

).

In this case, a retort cleaning cycle MUST be

carried out; this function cannot be canceled.

The water step can only be selected if cleaning

water is present in one of the bottles.

In "Auto by RMS" mode, "Cleaning water"

must have been defined for a station.

In "Concentration" mode, the process water is

automatically used as cleaning water.

If the magnetic stirrer was removed from the

retort, it must be reinserted before the start

of cleaning.

Otherwise, the cleaning reagents in the retort

will not be heated effectively.

If cassettes/baskets are not removed immediately after a paraf-
fin run, paraffin can drop into the previously emptied retorts.

After an infiltration program, it is not possible

to avoid cleaning - no CANCEL key is present.

Cleaning MUST be carried out!