5.1.5 The Service functions menu

• There is a button for each paraffin bath and for the paraffin station. The time

(in minutes) until the paraffin has melted enough so that it can be used in a
program step is shown next to each button.

• A program can already be started before that time – the program then extends

its delay step until the paraffin has melted enough to be used when the paraffin
step is reached.

• The time until the paraffin is fully melted as determined by the instrument

can be corrected by selecting the PARAFFIN BATH MELTING TIME button.
This function should normally only be used to increase the melting time, e.g.
if paraffin pellets have been added to a bath by hand.

• When the button is enabled, a window opens in which the value can be

changed.

Fig. 105

Paraffin bath melting times

Fig. 106

720

5. Operation

Certain settings and processes in the
ASP6025 can be changed using the
service functions.

Caution!

The functions that can be accessed in this

menu should only be used by experienced

users since incorrect use of these options can

lead to malfunctions or an impairment of the

process in the instrument.