1 the retort, Instrument setup – Leica Biosystems ASP6025 User Manual

Page 41

Advertising
background image

39

Leica ASP6025

4.4.1 The retort

4. Instrument Setup

Retort locking
• The retort locks automatically when a program is started.
• Unlock it by pressing on the unlocking mechanism switch

(8.

2

) below the screen (

Fig. 29

).

• A message box appears on the screen; YES must be entered

here to confirm unlocking of the retort while a process is in
progress.

• When the YES button is pressed (

Fig. 29

), the reagent vapors

are removed by suction, filtered and the fluid level in the
retort is lowered. Then the pin (43) unlocks the handle (8.

1

)

and the retort can be opened.

• Pressing NO continues the process.

Fig. 28

43

• To continue the process after opening the retort, move the

handle back into the locking position (

Fig. 28

). Then press

the CONTINUE button on the screen.

8.

1

• To open the retort, twist forward the black handle (8.

1

) on

the retort lid (4) (arrow in

Fig. 27

). The lid opens upwards.

Fig. 27

8.

1

4

Fig. 29

8.

2

Locked

Open

Pin (43) can also be unlocked manually - e.g.

after a power failure. To do so, push down the

pin using the Allen key provided.

Keep your distance when opening the retort

lid, particularly if the reagents are heated.

Also avoid breathing in the vapors.

Advertising