162
Instructions for Use V 2.2 RevI – 08/2013
6.
Cleaning and Maintenance
Fig. 231
You will be prompted to clean the sensors by the instrument
software at the end of each cleaning program (
Fig. 231
).
Proceed as follows:
• Remove the sieve (40,
Fig. 232
) in front of the sensors.
Cleaning the level sensors
• Use the cleaning tool together with the microfiber cloth as
shown in
Fig. 233
and
234
!
Fig. 232
Fig. 234
Fig. 233
40
The level sensors must be cleaned of all resi-
dues (reagents and paraffin) after every clean-
ing program.
NEVER use the cleaning tool WITHOUT the ac-
companying microfiber cloth, as otherwise the
sensors will be scratched!