Cleaning and Maintenance

Fig. 231

You will be prompted to clean the sensors by the instrument
software at the end of each cleaning program (

Fig. 231

).

Proceed as follows:
• Remove the sieve (40,

Fig. 232

) in front of the sensors.

Cleaning the level sensors

• Use the cleaning tool together with the microfiber cloth as

shown in

Fig. 233

and

234

!

Fig. 232

Fig. 234

Fig. 233

40

The level sensors must be cleaned of all resi-

dues (reagents and paraffin) after every clean-

ing program.

NEVER use the cleaning tool WITHOUT the ac-

companying microfiber cloth, as otherwise the

sensors will be scratched!