Operation, Tables with warning threshold values – Leica Biosystems ASP6025 User Manual

Page 101

Advertising
background image

99

Leica ASP6025

The RMS system – adjusting warning threshold values

(continued)

Tables with warning threshold values

5. Operation

Xylene protocols

Reagent

recommended replacement

Formalin

after 600 cassettes

70 % ethanol

after 600 cassettes

85 %/90 % ethanol

after 1500 cassettes

1 100 % ethanol

after 1500 cassettes

1 xylene

after 1500 cassettes

1 paraffin

after 1800 cassettes

Cleaning xylene

after 6 cycles

Cleaning alcohol

after 6 cycles

Cleaning water

after 6 cycles

Xylene-free protocols

Reagent

recommended replacement

Formalin

after 600 cassettes

70 % ethanol

after 600 cassettes

85 % ethanol

after 1500 cassettes

1 (ethanol/isopropanol) 80/20

after 1500 cassettes

1 isopropanol 100 %

after 1500 cassettes

ParaLast

after 2400 cassettes

1 paraffin

after 2400 cassettes

Cleaning xylene

after 6 cycles

Cleaning alcohol

after 6 cycles

Cleaning water

after 6 cycles

Advertising