2 basic instrument/hardware, 1 installation site requirements, Basic instrument/hardware – Leica Biosystems ASP6025 User Manual

Page 37: Instrument setup

Advertising
background image

35

Leica ASP6025

4. Instrument Setup

4.2 Basic instrument/hardware

Fig. 20

• The instrument requires an installation area of approx. 700 x 800 mm.
• The substrate must have a sufficient load capacity and rigidity with respect to the weight

of the instrument.

• Relative humidity 80 % maximum - non-condensing.
• Room temperature consistently between +15 °C and +40 °C.
• Elevation: up to a max. 2000 m above sea level.
• Ambient pressure from 740 hPa to 1100 hPa.
• The instrument is designed for indoor use only.
• The power supply must be within the range of the length of the power cable. No extension

cable may be connected.

• The instrument MUST be connected to a grounded power socket.
• Use only one of the provided power cables that is intended for the local power supply.
• Avoid vibrations, direct sunlight and heavy variation in temperature.

4.2.1 Installation site requirements

CAUTION!

The chemicals used in the Leica ASP6025 are easily inflammable and hazardous to health.

Therefore the installation location must be well ventilated, and must contain no open flames. The room in

which the instrument is located must not be used as a permanent station for people. Otherwise the room must

be equipped with an exhaust device.

The installation location must be protected against electrostatic discharge.

The instrument must be set up so that the power switch on the rear side of the instrument (

Fig. 3, item 37

) and

the power plug are easily accessible at all times.

Do not operate the instrument in rooms with explosion hazard.

To ensure proper function of the instrument, it must be set up while maintaining a minimum distance of 10 cm

from walls and furniture.

After unpacking the instrument, the instrument must be handled only by

the handles (26) on the rear side (

Fig. 3

) to move it to its final location. After-

wards, the brakes must be applied to the instrument's castors.

Advertising