5 the paraffin station, Initial operation – Leica Biosystems ASP6025 User Manual

Page 46

Advertising
background image

44

Instructions for Use V 2.2 RevI – 08/2013

Fig. 39

• The paraffin station must be pulled out to be filled. It has a cover (55) for

better thermal insulation and as splash protection. The cover has two holes
in it to make it easier to remove.

53

4.4.5 The paraffin station

• The paraffin station (54) is located behind the left door (53) of the instru-

ment, below the retort (

Fig. 37

). It ensures that sufficient fresh, liquid

paraffin is always available for the paraffin baths. The capacity is 5.0 l of
liquid paraffin.

• There are two marks in the interior of the station that display the minimum

fill level when filled with pellets or liquid paraffin (

Fig. 38

). The level must

not fall below these marks.

56

55

4.

Initial Operation

Fig. 37

56

Detail:

Sieve in the

paraffin station

54

Fig. 38

Top mark:

Minimum fill level when filling

with pellets for melting.

Bottom mark:

Minimum fill level when filling

with liquid paraffin.

Fig. 40

• The paraffin station can be filled with paraffin pellets or liquid paraffin.

If filled with pellets, the melting time is approx. 6.0 h.

• Check and clean the sieve (56) monthly.

If the paraffin station has been pulled out, this is indicated

by a red border (Fig. 40) on the SMART screen.

The indicator light on the lower right of the icon will also

be red.

Proceed slowly and carefully when pulling out the

station—never jerk it out. The paraffin inside it is fluid and

hot—it can cause burns. The cover is also hot, therefore

always wear gloves!