4 reagent handling, 1 filling/draining reagents, Reagent handling – Leica Biosystems ASP6025 User Manual

Page 115: Operation

Advertising
background image

113

Leica ASP6025

Fig. 155

Hose for remote filling/draining of reagents
• Connect the hose for remote filling/draining (70) as shown

in

Fig. 155

.

• When connecting the hose to the connection spout (item

18 in

Fig. 15

5) for reagents, the connection piece (71) must

engage with the spout with a clearly audible click.

18

18

70

71

Remote replacement of reagents

70

5.4 Reagent handling

5.4.1 Filling/draining reagents

5. Operation

The following steps should only be carried out

by trained laboratory personnel, experienced

in handling reagents.

Always wear rubber gloves and safety gog-

gles when handling the reagents used in this

instrument.

Reagents used for tissue infiltration can be

both toxic and/or flammable.

When filling or draining a system bottle or

RTU-bottle, ensure that the remote fill/drain

hose is securely hooked into the external sta-

tion and is not removed until the operation is

fully completed. Finally, pressurized air is used

to clear the hose after each fill / drain so that

no dirt remains. The hose must therefore not

be taken out of the external station until this

cleaning step is complete.

• The remote draining or filling function is possible in

3.8 and 5 liter mode. - Both system bottles and RTU-bottles
can be filled or drained.