5.1.6 System status

In the top right corner in the status bar, the general SYSTEM

STATUS is displayed via a small symbol.
Touching this symbol will lead you into the menu window of the
same name (

Fig. 11

3).

The status of the instrument is continuously monitored while the
ASP6025 is in operation. The messages of all sensors and active
moving parts (pumps, valves, etc.) are evaluated and used to
determine the system status.

System status

Fig. 109

For system status, there are three different levels. The different levels are displayed via a different
symbol for each in the top right corner of the title bar.
The SYSTEM STATUS symbol appears in all menus of the ASP6025 software and is constantly being
re-evaluated. If there is a system status change, then the corresponding new symbol is displayed.

Symbol

System status

OK

Warning

Error

Error status:
Serious errors of the ASP6025 have been detected during operation. The detected error is displayed
on the screen in a message. It must be confirmed by the user and has been included in the table
(

Fig. 113

). The program cannot be started.

Warning status:
There are some small problems, but none that would negatively affect infiltration. The exist-
ing problem was displayed in a message box on the screen and has been included in the table
(

Fig. 11

3). A reminder to this effect will appear again when the program is started.

Status OK:
The instrument works as expected and without any disruption.

Fig. 110

5. Operation

In the case of a "WARNING" status, a screen message is output when an infil-

tration program is started. The user can decide whether the program should

be started.

However, it is possible to start an infiltration program.

If the system status is "ERROR", it is not possible to start an infiltration

program.