116

Instructions for Use V 2.2 RevI – 08/2013

Remote filling

Select station

Press button

Fig. 162

Draining/filling reagents

(continued)

Fig. 163

Fig. 164

• After you have selected a reagent, a screen message appears

prompting you to make the necessary preparations.

• When everything is finished, press YES to start filling

(

Fig. 164

).

5. Operation

• On the screen, select an empty system bottle (e.g. S6) and

the external receptacle (Remote Station,

Fig. 162

).

• Then press the STATION REFILL button.
• Now the reagent with which the highlighted bottle is to

be filled (of course, the reagent contained in the external
station) must be selected (

Fig. 163

).

Remote filling of system bottles takes place in exactly the same
way as remote draining of RTU-bottles, but in the reverse order
of course.

+

Remote filling is only possible for system bot-

tles. RTU-bottles in the drawer must be filled

manually.