Operation – Leica Biosystems ASP6025 User Manual
Page 118

116
Instructions for Use V 2.2 RevI – 08/2013
Remote filling
Select station
Press button
Fig. 162
Draining/filling reagents
(continued)
Fig. 163
Fig. 164
• After you have selected a reagent, a screen message appears
prompting you to make the necessary preparations.
• When everything is finished, press YES to start filling
(
Fig. 164
).
5. Operation
• On the screen, select an empty system bottle (e.g. S6) and
the external receptacle (Remote Station,
Fig. 162
).
• Then press the STATION REFILL button.
• Now the reagent with which the highlighted bottle is to
be filled (of course, the reagent contained in the external
station) must be selected (
Fig. 163
).
Remote filling of system bottles takes place in exactly the same
way as remote draining of RTU-bottles, but in the reverse order
of course.
+
Remote filling is only possible for system bot-
tles. RTU-bottles in the drawer must be filled
manually.