Remote

Set-Up


AX3000 - User's Manual

191

When the set-up parameters have been loaded, the message '

OK,

set-up

sent

' is displayed on the operator’s console. If other messages are displayed,

refer to chapter 10.2.3.

The configuration file lists each set-up parameter. Refer to Appendix A.5 for a
detailed description of this file.

10.2.3 - Error Messages

There are two types of error:

- fatal errors: numbered from 100 to 114,
- warning errors: numbered from 200 to 221.

a) Fatal Errors

Fatal errors stop the remote command. The following fatal errors may occur:


ERR 101: header label expected.

Command:

setup_send

Cause:

the AXEL header label (BEGIN_AX_SETUP) is not the first
line of the configuration file.

Consequence: the AX3000 set-up is not modified.
Solution:

check the file. Replace it with a correct configuration file.


ERR 102: bad header label protocol version.

Command:

setup_send

Cause:

the protocol version (included in the header label) cannot be
processed by the current AX3000 firmware.

Consequence: the AX3000 set-up is not modified.
Solution:

use a configuration file with a compatible protocol.