Scream! 4.5

open with a text entry box. This window lets you edit Scream!'s

calibration values file.

Fill in the text box with calibration information for your digitiser and
the instrument connected to it, in the format described below.

To set the serial number of the instrument, include the line

Serial-Nos=serial-number

Scream! cannot tell what instrument is connected to the

digitiser. This line is provided to help you remember which set
of calibration values you have used, and to provide a title for

calibration graphs. If you attach a different instrument to the
same digitiser, you will need to enter new calibration values to

reflect the new instrument.

To set the sensitivity of the digitiser, include the line

VPC=sensitivity

VPC stands for voltage per count, measured in units of V/co

μ

unt.

This is sometimes given as V/Bit

μ

on the digitiser calibration

sheet.

To set the sensitivity of the calibration channel, include the line

CALVPC=sensitivity

as for the other digitiser channels.