User guide

When you have filled in all the values, click

.

Any WaveView windows that are open will change to show streams in

physical units. New WaveView windows will also use these units
where possible.

Each digitiser System ID and serial number can have only one

instrument connected to it. If you have a 6-channel digitiser with two
connected sensors, you will need to make the digitiser announce

different serial numbers for each one. On newer Güralp Systems
DM24 digitisers, this can be done with the command SERIAL2. See

the manual for your digitiser for more information.

3.8.1 Examples

The calibration information for a CMG-3T weak-motion velocity

sensor might look like the following:

Serial-Nos=T3X99

VPC=3.153,3.147,3.159

G=1010,1007,1002

COILCONST=0.02575,0.01778,0.01774

CALVPC=3.161

CALRES=51000

TYPE=CMG-3T

RESPONSE=CMG-3_30S_50HZ V

GRAVITY=9.80122

CMG-5TD accelerometers use 1 Ω calibration resistors, and their coil
constant is set to unity. Older CMG-5TD instruments, based on Mk2

digitiser hardware, do not have calibration input facilities, and thus
the CALVPC entry is omitted. For example: