Scream! 4.5

Sensor

Sensor type code

Units

(V/A)

CMG-3T, 360 s – 50 Hz response

CMG-3_360S_50HZ

V

CMG-3TB or 3V / 3ESP borehole,
30 s – 50 Hz response

CMG-3B_30S_50HZ

V

CMG-3TB or 3V / 3ESP borehole,
100 s – 50 Hz response

CMG-3B_100S_50HZ

V

CMG-3TB or 3V / 3ESP borehole,
120 s – 50 Hz response

CMG-3B_120S_50HZ

V

CMG-3TB or 3V / 3ESP borehole,
360 s – 50 Hz response

CMG-3B_360S_50HZ

V

CMG-3TB or 3V / 3ESP borehole,
360 s – 100 Hz response

CMG-3B_360S_100HZ

V

Some English descriptions are also accepted, e.g. “120s
velocity”, “100Hz acceleration” but this is not a free-format,
parsed field.

To set the sensitivity (or gain) of the sensor components,
include the line

G=vertical-sensitivity,N/S-sensitivity,E/W-sensitivity

These values are given on the sensor calibration sheet. For
velocity sensors, they are given in units of V m

-1

s (V/ms

-1

). The

gain of an accelerometer is expressed in V m

-1

s

2

(V/ms

-2

).

Because Güralp Systems sensors and digitisers use differential

inputs and outputs, the sensitivity is quoted as
2 × (single-ended sensitivity) on the calibration sheet.

To set the coil constants of the sensor components, include the

line

COILCONST=vertical-constant,N/S-constant,E/W-constant

These values are given on the sensor calibration sheet.

To set the local acceleration due to gravity, include the line

GRAVITY=acceleration

You should give this value in ms

-2

, if you know it. If you miss

out this line, Scream! will use a standard average g value of

9.80665 ms

-2

.