Guralp Systems Scream User Manual

Page 23

Advertising
background image

User guide

To set the value of the calibration resistor, include the line

CALRES=resistance

Güralp Systems digitisers normally use a 51 kΩ resistor

(CALRES=51000).

To set the sensor type, include the line

TYPE=model-number

e.g. 3T, 5T, etc..

To set the response of the sensor, include the line

RESPONSE=response-type unit

Some of the values you can use are given in the table below. If
none of these match the response of your instrument, please

contact support for advice.

Sensor

Sensor type code

Units

(V/A)

CMG-5T or 5TD,
DC – 100 Hz response

CMG-5_100HZ

A

CMG-40T-1 or 6T-1,
1 s – 100 Hz response

CMG-40_1S_100HZ

V

CMG-40T-1 or 6T-1,
2 s – 100 Hz response

CMG-40_2S_100HZ

V

CMG-40T-1 or 6T-1,
10 s – 100 Hz response

CMG-40_10S_100HZ

V

CMG-40T, 20 s – 50 Hz response

CMG-40_20S_50HZ

V

CMG-40T, 30 s – 50 Hz response

CMG-40_30S_50HZ

V

CMG-3T or 3ESP,
30 s – 50 Hz response

CMG-3_30S_50HZ

V

CMG-40T, 60 s – 50 Hz response

CMG-40_60S_50HZ

V

CMG-3T or 3ESP,

60 s – 50 Hz response

CMG-3_60S_50HZ

V

CMG-3T or 3ESP,

100 s – 50 Hz response

CMG-3_100S_50HZ

V

CMG-3T or 3ESP,

120 s – 50 Hz response

CMG-3_120S_50HZ

V

January 2014

23