Device Replacement
The parameters of Device A that have been set by using the offline function can be downloaded
to Device B of the same model.
App.B22E.ai
Figure App.-B-22 Device Replacement Image
App.B23E.ai
Connect the device A to FildMate
Start DTM
Display Offline parameters
Set the value to parameters
Upload the parameter from device
Export to file
Exit DTM
Existing (Old) device A Operation
Replacing (New) device B Operation
Exit DTM
Connect the device B to FieldMate
Start DTM
Display Offline parameters
Import from file
Download to the device
Figure App.-B-23 Operation Flow of Device Setup