<E Specifications Common to FieldMate Basic and Advance>

E-10

IM 01R01A01-01E

9th Edition : May 31, 2013-00

E-1-3 Configuration Function

The applications of FieldMate functions overlap. The following usage is recommended for

device configuration.

Table E-1-6

Device Communication Protocol

Application

Recommended Function

HART

Status confirmation

Confirm an overview with Segment Viewer

and then confirm details with Device Viewer

Settings/adjustments

Device DTM

Device replacement

Parameter Manager

FOUNDATION fieldbus H1

Status confirmation

Confirm an overview with Segment Viewer

and then confirm details with Device Viewer.

Settings/adjustments

Device DTM when there is a device DTM; DD

Menu when there is no device DTM

Device replacement

Parameter Manager

PROFIBUS

BRAIN

ISA100

Status confirmation

Confirm an overview with Segment Viewer

Settings/adjustments

Device replacement

Device DTM

Advertising