Removing the high voltage/ charger assembly, Removing the high voltage/charger assembly – ZOLL M Series Defibrillator Rev R User Manual

Page 101

Advertising
background image

M Series Service Manual

91

6. Removing

the High
Voltage/
Charger
Assembly

Tools Required

• #2 Phillips screwdriver

• Orange stick

• Small needle nose pliers

Setup

1. Remove the Front Panel Assembly.

2. Remove the Upper Housing Assembly.

3. Remove the System Board.

4. Remove the Battery Interconnect Board.

To reinstall the High Voltage Charger Assembly,
reverse the steps.

WARNING! This unit may contain lethal
voltages. You MUST completely discharge the
high voltage capacitor before removing from
unit.

DO NOT SHORT THE TERMINAL ENDS OF THE CAPACITOR. Refer to Section 8, Step 2.