M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 58

Advertising
background image

M Series Service Manual

48

4

These values only apply for test set-ups utilizing the BP Pump 2 Simulator. Variations of the test set-up or different simulators may

produce readings outside the provided values and will require end-user facility to establish the appropriate offset and tolerances.
Please contact ZOLL Technical Support i f you require assistance establishing the offset o f y o u r simulator and test set-up.

Do this...

Verify that...

Pass/
Fail

22.1

Turn the selector switch to
MONITOR mode.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

22.2

Ensure that the LEADS
parameters is set to PADS
(default).

If necessary, press the LEADS
button to cycle through the values
to select PADS.

The M Series unit displays PADS in the Lead selection
field on the monitor.

o

o

22.3

Press the NIBP button on the
E Series front panel.

The M Series displays the following measurements:

Systolic pressure (115 - 125 mmHg)

Diastolic pressure (75 - 85 mmHg)

Mean pressure (88 - 98 mmHg).

o

o

22.4

Press the SUMMARY softkey,
then press the TREND softkey,
followed by the NIBP Trend
softkey.

The M Series unit displays a summary of the NIBP
measurements, including the pulse rate reading (in the
range of 77 - 83 bpm).

o

o

Record your results on the Maintenance Tests Checklist.