M Series Service Manual

5

1.0 Physical Inspection of the Unit

Tools Needed

Battery.

Test Setup

None.

Observe this...

Pass/Fail/NA

1.1

Housing

Is the unit clean and undamaged?

o o

1.2

Does the unit show signs of excessive wear?

o o

1.3

Does the handle work properly?

o o

1.4

Does the recorder door open and close properly?

o o

1.5

Are input connectors clean and undamaged?

o o

1.6

Are there any cracks in the housing?

o o

1.7

Do the front panel or selector switches have any damage or cracks?

o o

1.8

Are there any loose housing parts?

o o

1.9

Cables

Are all cables free of cracks, cuts, exposed or broken wires?

o o

1.10

Are all cable bend/strain reliefs undamaged and free of excessive cable
wear?

o o

1.11

Paddles

Do the adult and pedi plates have major scratches or show signs of damage?

o o o

1.12

Do the adult shoes slide on and off easily to expose the covered pedi plates?

o o o

1.13

Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)

o o o