Removing the system interconn ect board, Removing the system interconnect board – ZOLL M Series Defibrillator Rev R User Manual

Page 108

Advertising
background image

M Series Service Manual

98

9. Removing

the
System
Interconn
ect Board

Tools Required

• #2 Phillips screwdriver

Setup

1. Remove the Front Panel Assembly.

2. Remove the Upper Housing Assembly.

3. Remove the System Board.

4. Remove the Battery Interconnect Board.

5. Remove the High Voltage/Charger module.

6. Remove the clear plastic print head isolater.

7. Remove the motor cable, cable print head

and paper sensor cable.

To reinstall the System Interconnect Board:

1. Install the laminate cables and keepers.

2. Ensure that the PCMCIA slots with gaskets

are located to the right rear of the board and
seated properly.

3. Install two screws.

4. Reconnect all the cables and reverse the steps.