M Series Service Manual
98
9. Removing
the
System
Interconn
ect Board
Tools Required
• #2 Phillips screwdriver
Setup
1. Remove the Front Panel Assembly.
2. Remove the Upper Housing Assembly.
3. Remove the System Board.
4. Remove the Battery Interconnect Board.
5. Remove the High Voltage/Charger module.
6. Remove the clear plastic print head isolater.
7. Remove the motor cable, cable print head
and paper sensor cable.
To reinstall the System Interconnect Board:
1. Install the laminate cables and keepers.
2. Ensure that the PCMCIA slots with gaskets
are located to the right rear of the board and
seated properly.
3. Install two screws.
4. Reconnect all the cables and reverse the steps.