M Series Service Manual

33

17.0 SpO

2

Monitor Test for SpO

2

Option

Tools Needed:

• Masimo



Reusable Sensor.

• Masimo Patient Cable

• Fluke Index Series 2 SpO

2

Simulator.

Test Setup:

Connect the universal cable to the MFC test plug.

DO NOT connect the ECG cable to the simulator.

Connect the Masimo Patient Cable and attach the Masimo sensor to the patient cable.

Connect the Masimo sensor to the finger simulation post.

Place a fully charged battery into the battery well or connect to ac power (dc power, if equipped).

Ensure that the SpO

2

Simulator is off.

Do this...

Verify that...

Pass/
Fail

17.1

Turn the selector switch to
MONITOR.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

The SpO

2

saturation percentage appears as a dashed

line on the monitor.

o

o

17.2

Wait ten seconds.

Turn on the SpO2 simulator.

Press the SIM softkey on the
Index SpO

2

Simulator, then press

the MAN softkey.

The unit displays the SpO

2

PULSE SEARCH message.

o

o

17.3

Press the 02+ or 02- softkey of
the simulator until the SpO

2

output is at 98%.

The SpO

2

reading of 98 +/- 1% appears on the E Series

monitor.

Note:

You may need to wait up to 2 minutes for the
information to appear on the ZOLL display.

o

o