M Series Service Manual

7

2.0 Front Panel Button Test

Tools Needed

QED 6 Defibrillator Analyzer

Test Setup

Install strip chart paper into the recorder tray.

Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.

Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the QED 6 Analyzer (or equivalent).

Do this...

Verify that...

Pass/Fail/
NA

2.1

Turn the selector switch to
MONITOR.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

The unit sounds 4 beep tones. PADS and MONITOR
display on the monitor.

Note:

PADS is a factory default setting.

o o

2.2

Set the QED 6 analyzer to NSR
of 120 BPM. To check the size of
the ECG waveform, press the
SIZE button.

As you press the SIZE button five times (0.5, 1.0, 1.5,
2.0, 3.0), the size of the ECG waveform appropriately
changes on the display.

o o

2.3

Press the ALARM SUSPEND
button.

Alarm symbol changes from disabled to enabled. If the
alarm sounds, press the ALARM SUSPEND button to
turn it off. The alarm will only be suspended for 90
seconds at this point. Press and hold the ALARM
SUSPEND
button for 3 seconds to disable alarms.

o

o

o

2.4

Press the RECORDER button.

The strip chart paper moves out of the unit from the
paper tray. Check that the correct time, date, ECG
lead annotation and waveform are recorded on the
paper. (Set Time and Date, if necessary.)

o

o

o

2.5

Open the paper compartment
door, then press the
RECORDER button.

The CHECK RECORDER message appears on the
monitor.

o

o

o

2.6

Close the paper compartment
door, then press the
RECORDER button.

The strip chart paper flows out of paper tray and the
monitor no longer displays the CHECK RECORDER
message.

o

o

o