M Series Service Manual

4

Equipment You Need for the M Series Options Maintenance

Tests

• Fluke Biomedical

Index 2 SpO

2

Simulator or equivalent. (For SpO

2

units only.)

• SpO

2

cable

and sensor (if option is installed).

• EtCO

2

cable,

and

CAPNOSTAT 3 Mainstream cable with airway adapter, or CAPNOSTAT 3 Sidestream cable with

cannula (if option is installed).

• Novametrix Capnostat Simulator Yb 1265/7100

• Paddles (if used).

• )

• Fluke Biomedical BP Pump 2 NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP

option is installed), or

• Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)

• IBP Temperature Simulator 9100-0402-TF

• DNI Nevada 214B Patient SImulator, with M Series interface cable

NOTE The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer
use different technologies for testing NIBP monitors and therefore, the manual provides two different procedures for
performing the NIBP Volume Leak test with each of these types of test equipment.