M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 19

Advertising
background image

M Series Service Manual

9

2.17

Remove the battery.

The charger light alternately flashes green and amber.

o

o

2.18

Replace the battery and the turn
unit on.

The amber charger light illuminates.

o

o

2.19

Press the ANALYZE button.

The SELECT DEFIB MODE message appears on the
monitor (for manual devices).

o

o

o

2.20

Connect the universal cable to
the QED 6. Move the selector
switch to DEFIB. Select 2J.
Press the CHARGE button.

The display shows that the unit is charging. The
SHOCK button lights when the unit is charged, and
the Ready tone for DEFIB sounds.

o

o

o

2.21

Press and hold the ENERGY
SELECT
down arrow.

The unit discharges internally and selected energy
decrements to 1J.

o

o

o

2.22

Press and release the ENERGY
SELECT
up arrow 19 times.

The following selections appear: Verify the following
energy amounts display incrementally:

Biphasic:2-10, 15, 20,30, 50, 75, 100,120, 150, 200J

Monophasic(DSW): 2-10, 15,20,30,50,75, 100, 150,
200, 300, 360J

o

o

o

2.23

Press the CHARGE button.

The display shows the unit charged up to 200J (360J
monophasic) and the SHOCK button lights.

o

o

o

2.24

Press the SHOCK button.

The unit discharges and the SHOCK button is no
longer illuminated.
A 15 second strip chart automatically prints, displaying
the number of joules delivered (if equipped with
recorder and configured to print event).

o

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Verify that...

Pass/Fail/
NA